Published April 2016 | Version v1
Journal article

Determination of optimal number of beams in direct machine parameter optimization-based intensity modulated radiotherapy for head and neck cases

  • 1. Philips Radiation Oncology Systems, Philips Ltd., Bengaluru (India)
  • 2. Department of Radiotherapy, Sanjay Gandhi Post Graduate Institute of Medical Sciences, Lucknow (India)

Description

This paper aims to introduce an algorithm called and 8220; sensitivity-based beam number selection (SBBNS)" for fully automated and case-specific determination of an optimal number of equispaced beams in intensity-modulated radiotherapy (IMRT). We tested the algorithm in five head and neck cases of varying complexity. We used direct machine parameter optimization method coupled with Auto Plan feature available in Pinnacle TPS (Version 9.10.0) for optimization. The Pearson correlation test shows a correlation of 0.88 between predicted and actual optimal number of beams, which indicates that SBBNS method is capable of predicting optimal number of beams for head and neck cases with reasonable accuracy. The major advantage of the algorithm is that it intrinsically takes into account various case- and machine-specific factors for the determination of optimal number. The study demonstrates that the algorithm can be effectively applied to IMRT scenarios to determine case specific and optimal number of beams for head and neck cases. (author)

Additional details

Publishing Information

Journal Title
Journal of Medical Physics
Journal Volume
41
Journal Issue
2
Journal Page Range
p. 129-134
ISSN
0971-6203

INIS

Country of Publication
India
Country of Input or Organization
India
INIS RN
51080931
Subject category
S62: RADIOLOGY AND NUCLEAR MEDICINE;
Descriptors DEI
ACCURACY; BEAMS; EFFECTIVE RADIATION DOSES; HEAD; NECK; RADIOTHERAPY
Descriptors DEC