Published October 1994 | Version v1
Report Open

Characterization inconsistencies in CdTe and CZT gamma-ray detectors

Description

In the past few years, significant developments in cadmium telluride (CdTe) and cadmium zinc telluride (CZT) semiconductor materials have taken place with respect to both quality and yield. Many of the more recent developments have occurred in the area of CZT crystal growth. This has resulted in an explosion of interest in the use of these materials in ambient temperature gamma-ray detectors. Most, if not all, of the manufacturers of CdTe and CZT have acquired government funding to continue research in development and applications, indicating the importance of these improvements in material quality. We have examined many detectors, along with the accompanying manufacturer's data, and it has become apparent that a clear standard does not exist by which each manufacturer characterizes the performance of their material. Result is a wide variety of performance claims that have no basis for comparison and normally cannot be readily reproduced. This paper first supports our observations and then proposes a standard that all manufacturers and users of these materials may use for characterization

Availability note (English)

MF available from INIS under the Report Number; Also available from OSTI as DE95009567; NTIS; US Govt. Printing Office Dep.

Files

26060422.pdf

Files (345.0 kB)

Name Size Download all
md5:b74cbe5057646d34fa1fd7a8889a2b9f
345.0 kB Preview Download

Additional details

Publishing Information

Imprint Pagination
5 p.
Report number
UCRL-JC--117142

Conference

Title
medical imaging conference.
Acronym
Nuclear science symposium
Dates
30 Oct - 5 Nov 1994.
Place
Norfolk, VA (United States).

INIS

Country of Publication
United States
Country of Input or Organization
United States
INIS RN
26060422
Subject category
S46: INSTRUMENTATION RELATED TO NUCLEAR SCIENCE AND TECHNOLOGY;
Resource subtype / Literary indicator
Conference
Descriptors DEI
CADMIUM TELLURIDES; CDTE SEMICONDUCTOR DETECTORS; GAMMA DETECTION; PERFORMANCE; SEMICONDUCTOR DETECTORS; STANDARDS; ZINC TELLURIDES
Descriptors DEC
CADMIUM COMPOUNDS; CHALCOGENIDES; DETECTION; MEASURING INSTRUMENTS; RADIATION DETECTION; RADIATION DETECTORS; TELLURIDES; TELLURIUM COMPOUNDS; ZINC COMPOUNDS

Optional Information

Contract/Grant/Project number
Contract W-7405-ENG-48
Funding organization
USDOE, Washington, DC (United States).
Secondary number(s)
CONF-941061--18.