Published March 1994
| Version v1
Journal article
Implementation of Tuy's cone-beam inversion formula
Description
Tuy's cone-beam inversion formula was modified to develop a cone-beam reconstruction algorithm. The algorithm was implemented for a cone-beam vertex orbit consisting of a circle and two orthogonal lines. This orbit geometry satisfies the cone-beam data sufficiency condition and is easy to implement on commercial single photon emission computed tomography (SPECT) systems. The algorithm which consists of two derivative steps, one rebinning step, and one three-dimensional backprojection step, was verified by computer simulations and by reconstructing physical phantom data collected on a clinical SPECT system. The proposed algorithm gives equivalent results and is as efficient as other analytical cone-beam reconstruction algorithms. (Author)
Additional details
Publishing Information
- Journal Title
- Physics in Medicine and Biology
- Journal Volume
- 39
- Journal Issue
- 3
- Journal Page Range
- p. 493-507.
- ISSN
- 0031-9155
- CODEN
- PHMBA7
Conference
- Title
- 2. international meeting on fully three-dimensional image reconstruction in radiology and nuclear medicine.
- Dates
- 23-25 Jun 1993.
- Place
- Snowbird, UT (United States).
INIS
- Country of Publication
- United Kingdom
- Country of Input or Organization
- United Kingdom
- INIS RN
- 25049800
- Subject category
- S62: RADIOLOGY AND NUCLEAR MEDICINE;
- Resource subtype / Literary indicator
- Conference
- Descriptors DEI
- ALGORITHMS; IMAGE PROCESSING; PHANTOMS; SINGLE PHOTON EMISSION COMPUTED TOMOGRAPHY; SPATIAL RESOLUTION
- Descriptors DEC
- COMPUTERIZED TOMOGRAPHY; EMISSION COMPUTED TOMOGRAPHY; MOCKUP; RESOLUTION; STRUCTURAL MODELS; TOMOGRAPHY