Published May 1985 | Version v1
Journal article

3 MeV electron beam radiotherapy

  • 1. Akita Univ. (Japan)

Description

Fourteen patients were treated with 3 MeV electron beam by a linear accelerator ''Mevatron 77'' from Apr. 1, 1982 to date. The depth dose curve and distribution of 3 MeV electron beams were quite stable and uniform, whether the field size was quite small as for the treatment of ilidocyclitis of ALL patient or large enough as for total body skin irradiation of mycosis fungoides. Dosimetry of 3 MeV electron beams revealed the electron build up of 7 mm in water, so that if the lesion was beneath the skin (2-3 mm, for example), suituable bolus material could be used. (author)

Additional details

Publishing Information

Journal Title
Hoshasen Chiryo Shisutemu Kenkyu
Journal Volume
2
Journal Issue
1
Series
Hoshasen Chiryo Shisutemu Kenkyu.
Journal Page Range
115-125
CODEN
HCSKE