Proximal surface caries detection with direct-exposure and rare earth screen/film imaging
Creators
- 1. Oregon Health Sciences Univ. School of Dentistry, Portland (USA)
Description
This laboratory study compared five imaging systems for their diagnostic accuracy in detection of proximal surface dental caries. Ten viewers provided data on radiographic detectability of carious lesions. The diagnostic accuracy of each system was determined with receiver operating characteristic (ROC) curves by comparing viewer data with the true state of the teeth as determined microscopically. D-speed film marginally outperformed the other four systems, but the three screen/film systems matched the diagnostic accuracy of E-speed film. Radiation reductions between 62% and 92% were achieved with the screen/film systems when compared to the two conventional dental films. The feasibility of designing a screen/film bite-wing cassette was shown, but the poor diagnostic accuracy of the present bite-wing system indicated a need for a new technology in caries detection
Additional details
Publishing Information
- Journal Title
- Oral Surgery, Oral Medicine, Oral Pathology
- Journal Volume
- 66
- Journal Issue
- 6
- Series
- Oral Surg., Oral Med., Oral Pathol.
- Journal Page Range
- 734-745
- ISSN
- 0030-4220
- CODEN
- OSOMA
INIS
- Country of Publication
- United States
- Country of Input or Organization
- United States
- INIS RN
- 20034012
- Subject category
- S62: RADIOLOGY AND NUCLEAR MEDICINE;
- Descriptors DEI
- ACCURACY; BIOMEDICAL RADIOGRAPHY; CARIES; COMPARATIVE EVALUATIONS; DENTISTRY; DIAGNOSIS; IMAGE PROCESSING; RADIATION DOSES; TEETH; X RADIATION
- Descriptors DEC
- DIAGNOSTIC TECHNIQUES; DIGESTIVE SYSTEM; DISEASES; ELECTROMAGNETIC RADIATION; EVALUATION; IONIZING RADIATIONS; MEDICINE; ORAL CAVITY; PATHOLOGICAL CHANGES; RADIATIONS