Published November 2011 | Version v1
Journal article

Calculation analysis of the thickness of radiation shield for the RIA equipment IP10

  • 1. Pusat Rekayasa Perangkat Nuklir, Badan Tenaga Nuklir Nasional, Kawasan Puspiptek, Serpong, Tangerang (Indonesia)

Description

Calculation Analysis has been performed on the thickness of radiation shield for the design of the Radioimmunoassay (RIA) IP10 counters using five detectors arranged in parallel. The calculation is intended to ensure that the radiation on each detector does not influence each other. The radiation shield is made of lead. The calculation of lead thickness was based on the principle of the lead plates absorptive power toward the gamma ray of a certain energy. which is the function of linear absorption coefficient and the material thickness. Assuming the use of Iodium-125(I-125) source with an activity 10 µCi, and expecting an absorptive power of 95%, calculations showed that the required lead thickness is equal to 0,013 cm. Since lead is soft and its availability in the market is limited, lead plate of 2 mm thickness are used instead, so that counting result for the detectors do not influence each other. (author)

Additional details

Additional titles

Original title (Indonesian)
Analisis perhitungan ketebalan perisai radiasi perangkat RIA IP10

Publishing Information

Journal Title
Prima (Tangerang)
Journal Volume
VIII
Journal Issue
2
Journal Page Range
p. 94-98
ISSN
1411-0296

Optional Information

Notes
3 refs.; 2 figs.