Published March 2004 | Version v1
Report Open

Neutron zooming detector using electronic zooming TV readout system

  • 1. Japan Atomic Energy Research Inst., Tokai, Ibaraki (Japan). Tokai Research Establishment
  • 2. Hamamatsu Photonics K.K., Electron Tube Division, Hamamatsu, Shizuoka (Japan)

Description

The neutron zooming tube has been developed using electronic zooming TV readout system for an X-ray microscope. This system consists of the photocathode, the electro-magnetic coil assembly and the CCD TV camera etc. The photocathode consists of transmission type CsI layer in combination with neutron converters such as Gd foil. The system position sensitive resolution of 25 μm has been obtained using neutron converter. We have found that it is determined from the distance between the CsI layer and the neutron converter, then the limiting resolution could be improved. (author)

Availability note (English)

Available from INIS in electronic form; Also available from JAEA; URL: http://jolisf.tokai-sc.jaea.go.jp/pdf/res/JAERI-Research-2003-032.pdf

Files

36116815.pdf

Files (1.8 MB)

Name Size Download all
md5:42d8aca412f86689b24821f167a45454
1.8 MB Preview Download

Additional details

Publishing Information

Imprint Pagination
30 p.
Report number
JAERI-Research--2003-032

Optional Information

Notes
5 refs., 11 figs., 2 tabs.; This record replaces 35098081