Minimum Detectable Activity(MDA) for Analysis of Noble Gas in Atmosphere
Description
The MDA value has been estimated in radioactivity analysis of noble gas in atmosphere. It is proposed how to reduce MDA in ultra low background sample and introduced several type of MDA used in the field. The minimum MDA value for total xenon is 0.4 mBq/m3 and the range of MDA value is almost below 1 mBq/m3. This factor is lower than that value proposed from the company of Germany. Here the MDA of 85Kr is proposed and the minimum time is determined
Availability note (English)
Also available from KAERIFiles
42052918.pdf
Files
(1.2 MB)
| Name | Size | Download all |
|---|---|---|
|
md5:4e213e6de466b9fc5ec21d4e6b5e3afb
|
1.2 MB | Preview Download |
Additional details
Publishing Information
- Imprint Pagination
- 42 p.
- Report number
- KAERI/TR--4133/2010
INIS
- Country of Publication
- Korea, Republic of
- Country of Input or Organization
- Korea, Republic of
- INIS RN
- 42052918
- Subject category
- S54: ENVIRONMENTAL SCIENCES;
- Descriptors DEI
- ATMOSPHERES; DETECTION; GASES; RADIOACTIVITY; XENON
- Descriptors DEC
- ELEMENTS; FLUIDS; GASES; NONMETALS; RARE GASES
Optional Information
- Notes
- 12 refs, 12 figs, 8 tabs