Published October 1987 | Version v1
Journal article

Continuous-monitoring detector-system of arterial H215O concentration for positron-emission tomography

  • 1. Research Inst. of Brain and Blood Vessels, Akita (Japan)

Description

It is required to measure arterial H215O concentration (input function) as accurately as possible, as well as tissue activity, in positron emission computed tomography. For this purpose, a small detector-system equipped with continuous blood-sampling tube and plastic scintilator was developed, whereby blood is aspirated through the radical artery and arterial H215O concentration is continuously monitoring. In this paper, the composition and performance of this system are presented. In addition, mathematical methodology for correcting ''dispersion'' that causes errors of cerebral blood flow measurement, and a simple method for adjusting the time-shift of blood-sampling curves are described. (Namekawa, K.)

Additional details

Additional titles

Subtitle (English)
Construction of the system and correction for the dispersion and the time-shift

Publishing Information

Journal Title
Kaku Igaku
Journal Volume
24
Journal Issue
10
Series
Kaku Igaku.
Journal Page Range
1587-1594
ISSN
0022-7854
CODEN
KAIGB