Published 1983 | Version v1
Journal article

The orthogonal hole transmission pattern (O.H.T.P.): its use of qualitive quality control for a Gamma Camera

  • 1. Hopital Henri-Mondor, 94 - Creteil (France)
  • 2. Oris, CEN Saclay, 91 - Gif-sur-Yvette (France)

Description

The 7,200 holes test pattern 0.H.T.P. recommended by the B.R.H. and the W.H.O. to appreciate with the course of time the intrinsic spatial resolution, the spatial linearity and the homogeneity of a Gamma Camera was used to demonstrate that this lead test allows a better analysis of non uniformities by decoupling sensitivity and spatial resolution parameters

Abstract (French)

La mire a 7200 trous M.T.L. preconisee par le B.R.H. et l'O.M.S. pour apprecier longitudinalement la resolution spatiale intrinseque, la linearite spatiale et l'homogeneite d'une Gamma Camera fut utilisee pour demontrer que ce test en plomb permet de mieux analyser les defauts d'homogeneite en decouplant les parametres de sensibilite et de linearite spatiale

Additional details

Additional titles

Original title (French)
Fantome multitrous lineaires (M.T.L.). Son utilisation pour le controle qualitatif d'une Gamma Camera

Publishing Information

Journal Title
J. Biophys. Med. Nucl.
Journal Volume
7
Journal Issue
3
Series
J. Biophys. Med. Nucl.
Journal Page Range
147-150
ISSN
0399-0435

INIS

Country of Publication
France
Country of Input or Organization
France
INIS RN
15025733
Subject category
S46: INSTRUMENTATION RELATED TO NUCLEAR SCIENCE AND TECHNOLOGY;
Descriptors DEI
GAMMA CAMERAS; PHANTOMS; QUALITY CONTROL; SPATIAL RESOLUTION
Descriptors DEC
CAMERAS; CONTROL; MOCKUP; RESOLUTION; STRUCTURAL MODELS