Published 1983
| Version v1
Journal article
The orthogonal hole transmission pattern (O.H.T.P.): its use of qualitive quality control for a Gamma Camera
Creators
- 1. Hopital Henri-Mondor, 94 - Creteil (France)
- 2. Oris, CEN Saclay, 91 - Gif-sur-Yvette (France)
Description
The 7,200 holes test pattern 0.H.T.P. recommended by the B.R.H. and the W.H.O. to appreciate with the course of time the intrinsic spatial resolution, the spatial linearity and the homogeneity of a Gamma Camera was used to demonstrate that this lead test allows a better analysis of non uniformities by decoupling sensitivity and spatial resolution parameters
Abstract (French)
La mire a 7200 trous M.T.L. preconisee par le B.R.H. et l'O.M.S. pour apprecier longitudinalement la resolution spatiale intrinseque, la linearite spatiale et l'homogeneite d'une Gamma Camera fut utilisee pour demontrer que ce test en plomb permet de mieux analyser les defauts d'homogeneite en decouplant les parametres de sensibilite et de linearite spatialeAdditional details
Additional titles
- Original title (French)
- Fantome multitrous lineaires (M.T.L.). Son utilisation pour le controle qualitatif d'une Gamma Camera
Publishing Information
- Journal Title
- J. Biophys. Med. Nucl.
- Journal Volume
- 7
- Journal Issue
- 3
- Series
- J. Biophys. Med. Nucl.
- Journal Page Range
- 147-150
- ISSN
- 0399-0435
INIS
- Country of Publication
- France
- Country of Input or Organization
- France
- INIS RN
- 15025733
- Subject category
- S46: INSTRUMENTATION RELATED TO NUCLEAR SCIENCE AND TECHNOLOGY;
- Descriptors DEI
- GAMMA CAMERAS; PHANTOMS; QUALITY CONTROL; SPATIAL RESOLUTION
- Descriptors DEC
- CAMERAS; CONTROL; MOCKUP; RESOLUTION; STRUCTURAL MODELS