Dose mapping in gamma chamber using glass plate dosimeter

  • 1. Defence Laboratory, Jodhpur (India)

Description

The quality of radiation-processed (blood, animal, medical and R and D) material is mostly dependent on how accurately the radiation dose has been delivered to the material. Chemical (Fricke, FBX and Ceric-cerous), ESR and TL are few processes on which high dose (kGy range) dosimetry is based. Applications of general-purpose glasses have been reported to for its use as high dose gamma dosimetry using, TL, OD (Optical Density) and spectro-photometric techniques in recent past. The dose distributions in the irradiation volume of gamma chamber are not perfectly homogeneous and variations reported within ± 25%. If dose in-homogeneity within the chamber is not under one's control, at least it should be known for dosimetry correction and uncertainty evaluation in the irradiated samples. Authors have employed general-purpose transparent glass plate for evaluating the dose distribution in any close type gamma chamber and results are presented

Additional details

Publishing Information

Journal Title
Journal of Medical Physics
Journal Volume
32
Journal Issue
suppl
Journal Page Range
p. S56
CODEN
JMPHFE

Conference

Title
28. annual conference of Association of Medical Physicists of India
Acronym
AMPICON-2007
Dates
2-4 Nov 2007
Place
Srinagar (India)

INIS

Country of Publication
India
Country of Input or Organization
India
INIS RN
40033538
Subject category
S61: RADIATION PROTECTION AND DOSIMETRY;
Resource subtype / Literary indicator
Conference
Descriptors DEI
CHEMICAL DOSEMETERS; DENSITOMETERS; GAMMA DOSIMETRY; RADIATION DOSE DISTRIBUTIONS; RADIATION DOSES
Descriptors DEC
DOSEMETERS; DOSES; DOSIMETRY; MEASURING INSTRUMENTS; PHOTOMETERS