Published January 2017 | Version v1
Journal article

Activity measurements of the radionuclides 18F and 99mTc for the NMISA, South Africa in the ongoing comparisons BIPM.RI(II)-K4.F-18 and BIPM.RI(II)-K4.Tc-99m

  • 1. BIPM - Bureau International des Poids et Mesures, F-92312 Sevres Cedex (France)
  • 2. National Metrology Institute of South Africa - NMISA, 0081 Pretoria Gauteng (South Africa)

Description

In 2015, comparisons of activity measurements of 18F and 99mTc using the Transfer Instrument of the International Reference System (SIRTI) took place at the National Metrology Institute of South Africa (NMISA, South Africa). Ampoules containing about 25 kBq of 18F and 99mTc solutions were measured in the SIRTI for more than two half-lives. The NMISA standardized the activity in the ampoules by ionization chamber measurements traceable to 4π(LS)β-γ coincidence measurements. The comparisons, identifiers BIPM.RI(II)-K4.F-18 and BIPM.RI(II)-K4.Tc-99m, are linked to the corresponding BIPM.RI(II)-K1.F-18 and BIPM.RI(II)-K1.Tc-99m comparisons and degrees of equivalence with the respective key comparison reference values have been evaluated. (authors)

Additional details

Publishing Information

Journal Title
Metrologia
Journal Volume
54
Journal Issue
no.1A
Journal Page Range
p. 06001.1-06001.22
ISSN
0026-1394

Optional Information

Notes
15 refs.