Published August 2013 | Version v1
Journal article

Influence of different frequencies and insertion depths on the diagnostic accuracy of liver elastography by acoustic radiation force impulse imaging (ARFI)

  • 1. Department of Gastroenterology, Hepatology and Endocrinology, Hannover Medical School, Carl-Neuberg-Str. 1, 30625 Hannover (Germany)

Description

Background: Acoustic Radiation Force Impulse Imaging (ARFI) is an innovative elastography for staging of liver fibrosis. We evaluated the diagnostic accuracy of different probes to perform ARFI at different insertion depths. Methods: In a prospective study, 89 chronic HCV infected patients underwent ARFI elastography using both available probes (c-ARFI: C4-1-MHz; l-ARFI: L9-4 MHz) in comparison to Fibroscan®. Variability of ARFI elastography at different insertion depths was systematically evaluated in 39 patients (44%). According to Fibroscan® elastography, 32 patients (36%) presented with liver cirrhosis, 23 patients (26%) had significant fibrosis and 34 patients (38%) had no significant fibrosis. Results: Mean propagation velocity with c-ARFI was 1.70 ± 0.67 m/s and 1.91 ± 0.87 m/s with l-ARFI. Results of both probes were correlated to each other (p < 0.001; r = 0.70) and to Fibroscan® (p < 0.001, r = 0.82 and 0.84, respectively). In patients with significant fibrosis or with cirrhosis, mean values by l-ARFI were significantly higher than by c-ARFI (p < 0.001). For detection of liver cirrhosis, AUROC was 0.97 for c-ARFI (cut-off level 1.72 m/s) and 0.90 for l-ARFI (cut-off 2.04 m/s). Correlation coefficients of c-ARFI with Fibroscan® were highest at an insertion depth of 5–6 cm (r = 0.882 and 0.864, respectively, p < 0.001) and at 3–4 cm for l-ARFI (r = 0.850 and 0.838, respectively, p < 0.001). Conclusions: ARFI elastography with the linear and with the convex probes showed comparable validity and accuracy in the estimation of liver stiffness. The linear probe gave higher ARFI values. The most accurate insertion depth was 5–6 cm for c-ARFI and 3–4 cm for l-ARFI indicating that measurements should not be performed close to the liver capsule

Availability note (English)

Available from http://dx.doi.org/10.1016/j.ejrad.2013.02.011

Additional details

Identifiers

DOI
10.1016/j.ejrad.2013.02.011;
PII
S0720-048X(13)00081-8;

Publishing Information

Journal Title
European Journal of Radiology
Journal Volume
82
Journal Issue
8
Journal Page Range
p. 1207-1212
ISSN
0720-048X
CODEN
EJRADR

INIS

Optional Information

Copyright
Copyright (c) 2013 Elsevier Science B.V., Amsterdam, The Netherlands, All rights reserved.