Published 1978 | Version v1
Report Restricted

Germanium detector system for the detection of transuranics at low-activity concentrations in soil

Description

A photon spectroscopy system is described which is designed for the detection of plutonium and 241Am in soil samples with a minimum turnaround time. Quantification is based upon the 60-keV gamma emitted in 241Am decay and upon the uranium L x-rays (energies from 13 to 22 keV) emitted by plutonium isotopes during alpha decay. The detector is a single-crystal, intrinsic-germanium-planar detector with a surface area of 21 cm2. Sensitivity is increased by incorporating a detector window with a larger than normal surface area. This optimized window size was established by Monte Carlo calculations. For small, Petri-dish samples, detection limits at the 3 sigma level for a 4-hr counting time are better than 4 pCi/g for plutonium and better than 0.05 pCi/g for 241Am. The specifications, performance, and cost of the system are discussed

Availability note (English)

MF available from INIS under the Report Number; Available from NTIS., PC A02/MF A01.

Files

Restricted

The record is publicly accessible, but files are restricted to users with access.

Additional details

Additional titles

Augmented title (English)
"2"4"1Am, "2"4"0Pu, "2"3"9Pu, "2"3"8U, "2"3"2Th

Publishing Information

Imprint Pagination
13 p.
Report number
LA-UR--78-87

Conference

Title
11. symposium on radiation instrumentation.
Dates
17 - 19 Jan 1978.
Place
San Diego, CA, USA.

Optional Information

Secondary number(s)
CONF-780110--10.