A simple quantitative method for assessment of multiple window spatial registration
Creators
- 1. Royal Perth Hospital, Perth, WA (Australia). Department of Medical Physics
Description
Full text: When acquiring images using multiple analyser windows, it is important that the image from each window registers in the same spatial location to prevent image degradation through loss of spatial resolution. The NEMA performance standard measurement for multiple window spatial registration (MWSR), while only requiring a simple jig, requires specialized analysis software that is not normally available. We have developed a test that uses a simple jig that provides a quantitative assessment of MWSR using software analysis tools that are available on most gamma camera systems. A 50 mm thick lead brick with a slot cut along its long axis and a 'V' groove in the top is used as a jib. A vial of activity laid in the top groove is collimated by the brick and produces a line source of activity. A 67Ga line source image, acquired using three analyser windows, and a 99Tcm line source image are collected using the jig for 5000 k counts each into a 1024 image matrix. Line profiles are drawn across both images and the full width at half maximum (FWHM) of each profile determined. Tools for performing this analysis are widely available on camera systems. The ratio of the 67Ca to 99Tcm FWHM values b used as a quantitative indication of the MWSR for the camera. An average ratio of 1.05 has been observed for a Toshiba 901A/HC gamma camera over a 2 year period. We have taken a ratio of 1.10 as an action threshold above which camera adjustment is required
Additional details
Publishing Information
- Journal Title
- Nuclear Medicine Communications
- Journal Volume
- 20
- Journal Issue
- 4
- Journal Page Range
- p. 385-386
- ISSN
- 0143-3636
Conference
- Title
- 30. Annual Scientific Meeting of the Australian and New Zealand Society of Nuclear Medicine
- Dates
- 24-27 Apr 1999
- Place
- Perth, WA (Australia)
INIS
- Country of Publication
- United Kingdom
- Country of Input or Organization
- Australia
- INIS RN
- 32008246
- Subject category
- S62: RADIOLOGY AND NUCLEAR MEDICINE;
- Resource subtype / Literary indicator
- Conference
- Descriptors DEI
- CALIBRATION STANDARDS; DATA ANALYSIS; GALLIUM 67; IMAGE PROCESSING; ISOTOPE RATIO; SCINTISCANNING; TECHNETIUM 99
- Descriptors DEC
- BETA DECAY RADIOISOTOPES; BETA-MINUS DECAY RADIOISOTOPES; COUNTING TECHNIQUES; DAYS LIVING RADIOISOTOPES; DIAGNOSTIC TECHNIQUES; ELECTRON CAPTURE RADIOISOTOPES; GALLIUM ISOTOPES; HOURS LIVING RADIOISOTOPES; INTERMEDIATE MASS NUCLEI; INTERNAL CONVERSION RADIOISOTOPES; ISOMERIC TRANSITION ISOTOPES; ISOTOPES; NUCLEI; ODD-EVEN NUCLEI; PROCESSING; RADIOISOTOPE SCANNING; RADIOISOTOPES; STANDARDS; TECHNETIUM ISOTOPES; YEARS LIVING RADIOISOTOPES
Optional Information
- Notes