Published October 7, 2012 | Version v1
Journal article

Calorimetric determination of the absorbed dose to water for medium-energy x-rays with generating voltages from 70 to 280 kV

  • 1. Physikalisch-Technische Bundesanstalt, Bundesallee 100, 38116 Braunschweig (Germany)

Description

For medium energy x-rays produced with tube voltages from 70 to 280 kV, the absorbed dose to water, Dw, has been determined by means of water calorimetry with relative standard uncertainties ranging from 0.45% to 0.98% at 280 and 70 kV. The results were confirmed by Monte Carlo calculations, in which the ratios of Dw at 5 cm depth in a reference water phantom to the air kerma free in air, Ka, at the same point in space were compared to the corresponding ratios determined experimentally. The general agreement between measurement and calculation was better than 1%. These results confirm earlier investigations in which the absorbed dose to graphite was determined by means of a graphite extrapolation chamber. For the Monte Carlo calculations, an attempt was made to present a complete uncertainty budget, taking into account type B contributions also. (paper)

Availability note (English)

Available from http://dx.doi.org/10.1088/0031-9155/57/19/6245

Additional details

Identifiers

Publishing Information

Journal Title
Physics in Medicine and Biology
Journal Volume
57
Journal Issue
19
Journal Page Range
p. 6245-6268
ISSN
0031-9155
CODEN
PHMBA7