Purpose: To evaluate the time required, the accuracy and the precision of a model-based image analysis software tool for the diagnosis of osteoporotic fractures using a 6-point morphometry protocol. Materials and methods: Lateral dorsal and lumbar radiographs were performed on 92 elderly women (mean age 69.2 ± 5.7 years). Institutional review board approval and patient informed consent were obtained for all subjects. The semi-automated and the manual correct annotations of 6-point placement were compared to calculate the time consumed and the accuracy of the software. Twenty test images were randomly selected and the data obtained by multiple perturbed initialisation points on the same image were compared to assess the precision of the system. Results: The time requirement data of the semi-automated system (420 ± 67 s) were statistically different (p < 0.05) from that of manual placement (900 ± 77 s). In the accuracy test, the mean reproducibility error for semi-automatic 6-point placement was 2.50 ± 0.72% [95% CI] for the anterior-posterior reference and 2.16 ± 0.5% [95% CI] for the superior-inferior reference. In the precision test the mean error resulted averaged over all vertebrae was 2.6 ± 1.3% in terms of vertebral width. Conclusions: The technique is time effective, accurate and precise and can, therefore, be recommended in large epidemiological studies and pharmaceutical trials for reporting of osteoporotic vertebral fractures.

Availability note (English)

Available from http://dx.doi.org/10.1016/j.ejrad.2007.12.001

Additional details

Identifiers

DOI
10.1016/j.ejrad.2007.12.001;
PII
S0720-048X(07)00602-X;

Publishing Information

Journal Title
European Journal of Radiology
Journal Volume
70
Journal Issue
1
Journal Page Range
p. 142-148
ISSN
0720-048X
CODEN
EJRADR

INIS

Country of Publication
Netherlands
Country of Input or Organization
International Atomic Energy Agency (IAEA)
INIS RN
41032919
Subject category
S62: RADIOLOGY AND NUCLEAR MEDICINE;
Descriptors DEI
ACCURACY; COMPUTER CODES; DIAGNOSIS; FRACTURES; IMAGE PROCESSING; IMAGES; OSTEOPOROSIS; RADIOLOGY; VERTEBRAE; WOMEN
Descriptors DEC
ANIMALS; BODY; DISEASES; FAILURES; FEMALES; MAMMALS; MAN; MEDICINE; NUCLEAR MEDICINE; ORGANS; PRIMATES; PROCESSING; SKELETAL DISEASES; SKELETON; VERTEBRATES

Optional Information

Copyright
Copyright (c) 2007 Elsevier Science B.V., Amsterdam, The Netherlands, All rights reserved.