An object-independent measure for improving misalignment correction in C-arm CT

  • 1. Erlangen-Nuernberg Univ., Erlangen (Germany). Inst. of Medical Physics (IMP)
  • 2. Siemens AG, Forchheim (Germany). Healthcare Sector

Description

Misalignment in c-arm CT is a frequently discussed problem. Numerous methods for misalignment correction were investigated, but up to now no appropriate measure for geometry correction algorithms has been given. We present a new object-in-dependent measure for geometry comparison in c-arm systems. Our approach is based on a virtual object and assigns one single value to compare geometry calibrations. Invariance towards variable calibration phantom positions as well as its independence towards different c-arm systems is demonstrated. We describe the characteristic behaviour for several causes of misalignment and the results of a simulated dataset are presented. (orig.)

Part of:
Fully three-dimensional image reconstruction in radiology and nuclear medicine. Proceedings

Additional details

Publishing Information

Imprint Title
Fully three-dimensional image reconstruction in radiology and nuclear medicine. Proceedings
Imprint Pagination
480 p.
Journal Page Range
p. 148-151

Conference

Title
11th international meeting on ''Fully three-dimensional image reconstruction in radiology and nuclear medicine'' and The 3rd workshop on ''High performance image reconstruction''
Dates
11-15 Jul 2011
Place
Potsdam (Germany)