Published 1996 | Version v1
Miscellaneous

Perturbation of 6MV photons by radiographic film: measurements and Monte Carlo simulations

  • 1. University of Sydney, Lidcombe, NSW (Australia). School of MRT
  • 2. Prince of Wales Hospital, Randwick, NSW (Australia). Department of Radiation Oncology
  • 3. Illawarra Cancer Care Centre, Wollongong, NSW (Australia). Department of Medical Physics

Description

Full text: Radiographic film is an established practical tool used in the measurement of dose distributions for radiotherapy purposes. The accuracy and reproducibility of film optical density as an indicator of dose, has been associated with several factors including photon energy, processing conditions and film plane orientation. Few studies have investigated the factors causing variability in film dosimetry, due to the difficulty of separating the individual contributions. The effect that a sheet of radiographic film in a water phantom has on the dose deposited by a 6MV photon beam when orientated perpendicular and parallel to the beam central axis is reported. These measurements are compared with Monte Carlo simulations. The Monte Carlo simulations were performed on a Silicon Graphics Indigo R4000, using the EGS4 code. Measurements were taken using a Varian Clinac 600 C 6MV beam incident on a solid water phantom, with a sheet of Kodak X Omat-V film. The effect of thickness of the film and its orientation with respect to the beam were investigated. Monte Carlo generated spectra were used to calculate collision kerma (KColl) for water and film elements at each depth. Results indicate an expected over response of 3.5% at 30 cm deep compared to 1.5 cm deep. The calculated results suggest an explanation in terms of the predominantly forward directed secondary electrons, for the measured difference in film response at depth for film parallel as compared to perpendicular to the central axis of the beam

Part of:
Engineering and physical sciences in medicine and health conference. Programme Book

Additional details

Publishing Information

Imprint Title
Engineering and physical sciences in medicine and health conference. Programme Book
Imprint Pagination
210 p.
Journal Page Range
p. 150

Conference

Title
EPSMH'96. Engineering and physical sciences in medicine and health conference
Dates
21-24 Oct 1996
Place
Canberra, ACT (Australia)

INIS

Country of Publication
Australia
Country of Input or Organization
Australia
INIS RN
33009281
Subject category
S61: RADIATION PROTECTION AND DOSIMETRY;
Resource subtype / Literary indicator
Conference, Non-conventional Literature
Descriptors DEI
ACCURACY; CALCULATION METHODS; FILM DOSIMETRY; KERMA; MEV RANGE 01-10; MONTE CARLO METHOD; PHANTOMS; PHOTON BEAMS; RADIATION DOSE DISTRIBUTIONS; RADIOTHERAPY; THICKNESS
Descriptors DEC
BEAMS; CALCULATION METHODS; DIMENSIONS; DOSIMETRY; ENERGY RANGE; MEDICINE; MEV RANGE; MOCKUP; NUCLEAR MEDICINE; RADIOLOGY; STRUCTURAL MODELS; THERAPY

Optional Information

Notes
This record replaces 30052620