Experimental determination of wall correction factors. Pt. 1

  • 1. The Netherlands Cancer Inst., Amsterdam (Netherlands). Radiotherapy Dept.

Description

Wall correction factors were measured for different types of ionisation chamber, both in air in a 60Co gamma-ray beam and in a water phantom irradiated by a 20 MeV electron beam. A comparison of the authors' values with other experimental values shows agreement, within 0.3%, after normalisation on reference chamber values, for most chambers. A 0.9% lower value was observed for a nylon-walled Farmer chamber suggesting that differences in thickness of the inner surface of the thimble might occur. Generally good agreement, within 0.5%, exists between experimental and calculated wall correction factors. For two chambers constructed from A-150 plastic larger deviations (0.6% and 1.4%) were observed. Due to these unexplained discrepancies the use of an average of calculated and measured values of wall correction factors in dosimetry protocols is recommended. (author)

Additional details

Additional titles

Subtitle (English)
Cylindrical ionisation chambers

Publishing Information

Journal Title
Physics in Medicine and Biology
Journal Volume
35
Journal Issue
7
Series
Phys. Med. Biol.
Journal Page Range
835-846
ISSN
0031-9155
CODEN
PHMBA