Published August 1981 | Version v1
Report Restricted

Detectors, sampling, shielding, and electronics for positron emission tomography

Description

A brief discussion of the important design elements for positron emission tomographs is presented. The conclusions are that the instrumentation can be improved by the use of larger numbers of small, efficient detectors closely packed in many rings, the development of new detector materials, and novel electronic designs to reduce the deadtime and increase maximum event rates

Availability note (English)

MF available from INIS under the Report Number; Available from NTIS, PC A02/MF A01 as DE81029858.

Files

Restricted

The record is publicly accessible, but files are restricted to users with access.

Additional details

Publishing Information

Imprint Pagination
9 p.
Report number
LBL--13091

Conference

Title
ANS winter meeting.
Dates
29 Nov - 4 Dec 1981.
Place
San Francisco, CA (USA).

INIS

Country of Publication
United States
Country of Input or Organization
United States
INIS RN
14772165
Subject category
S62: RADIOLOGY AND NUCLEAR MEDICINE;
Resource subtype / Literary indicator
Conference
Descriptors DEI
BIOLOGICAL SHIELDING; CONFIGURATION; EMISSION COMPUTED TOMOGRAPHY; EQUIPMENT; GAMMA CAMERAS; POSITRON COMPUTED TOMOGRAPHY; RADIATION DETECTORS; RECOMMENDATIONS; SPECIFICATIONS
Descriptors DEC
CAMERAS; COMPUTERIZED TOMOGRAPHY; DIAGNOSTIC TECHNIQUES; MEASURING INSTRUMENTS; SHIELDING; TOMOGRAPHY

Optional Information

Secondary number(s)