Comparison of transaxial resolution in 1800 and 3600 SPECT with a rotating scintillation camera

  • 1. Johns Hopkins Medical Institutions, Baltimore, MD (USA). Div. of Nuclear Medicine
  • 2. Johns Hopkins Medical Institutions, Baltimore, MD (USA). Div. of Radiation Health Sciences

Description

Using circular 1800 and 3600 SPECT acquisition modes the transaxial resolution of line sources in air and water were measured at different positions in the field of view. With the 1800 acquisition mode, all line sources in air located off the axis of camera rotation (AoR) showed an oval distortion. This distortion was systematically related to the starting point of the rotating detector. On axis line sources in air were undistorted, regardless of the 1800 acquisition starting angle. The 3600 acquisition images of the line sources in air showed a similar effect but in a very mild form. In water, transaxial reconstructions of line sources (off axis) showed an enhancement of the oval distortion for both the 1800 and 3600 acquisitions. Computer simulations of the line source measurements were performed and correlated well with the experimental data. The line source results are explainably by the inherent depth dependent response of the scintillation camera. In clinical SPECT studies, distortions of this nature will be most appreciable with 1800 imaging of small organs that are located off the AoR. (orig.)

Additional details

Publishing Information

Journal Title
Eur. J. Nucl. Med.
Journal Volume
13
Journal Issue
1
Series
Eur. J. Nucl. Med.
Journal Page Range
7-11
ISSN
0340-6997
CODEN
EJNMD

Optional Information

Contract/Grant/Project number
Grant CA 32845