A flow injection system with four ion exchange resin columns for cadmium pre-concentration and determination by flame AAS

  • 1. Sao Paulo Univ., Sao Carlos, SP (Brazil). Inst. de Quimica
  • 2. Centro de Energia Nuclear na Agricultura (CENA), Piracicaba, SP (Brazil)

Description

A flow injection system comprised of four mini-columns packed with cation exchange resin was designed to implement an on-line pre-concentration procedure for cadmium determination by flame atomic absorption spectrophotometry. The AG50W-X8 resin behavior in the flow system was studied, talking into account its suitability for application to routine analysis. A throughput of up to 80 determinations per hour, and a detection limit of 0.7 ng/L were achieved. The relative standard deviation of measurements (n=10) for a sample with 20.0 g Cd/L was lower than 2%. By applying the t-test at a probability level of 5% no significant was observed in comparing the results obtained by the proposed method and the certified values of standard reference materials. (author)

Additional details

Publishing Information

Journal Title
Journal of the Brazilian Chemical Society
Journal Volume
6
Journal Issue
4
Journal Page Range
p. 387-392
ISSN
0103-5053
CODEN
JOCSET

Optional Information

Notes
16 refs., 5 figs., 1 tab.