Doses on the central axes of narrow 6-MV x-ray beams
Creators
- 1. Joint Center for Radiation Therapy, Department of Radiation Therapy, Harvard Medical School, Boston, MA (USA)
Description
The absorbed doses on the central axes of narrow beams (radii 0.07--2.5 cm) of 6-MV x rays have been studied by experiments and Monte Carlo simulations. The measurements were made in a geometry used for irradiation of intracranial lesions. For radii <1.0 cm the dose on the central axis is progressively reduced due to electron disequilibrium. This leads to measurement artifacts when the detector is too large, as was readily observed with ionization chambers. Radiographic and radiochromic films were used with densitometric evaluation to provide the resolution necessary to measure absorbed doses for the narrowest beams. The contribution by phantom-scattered photons is significant even at small field sizes, and scatter factors were determined from the experimental results. Photons scattered by the auxiliary collimator did not add appreciably to the dose on the central axis. The data were used to characterize the dose-to-kerma ratio as a function of beam radius. Differences between experimental results and those from Monte Carlo calculations were observed
Additional details
Publishing Information
- Journal Title
- Medical Physics
- Journal Volume
- 17
- Journal Issue
- 5
- Series
- Med. Phys.
- Journal Page Range
- 794-799
- ISSN
- 0094-2405
- CODEN
- MPHYA
INIS
- Country of Publication
- United States
- Country of Input or Organization
- United States
- INIS RN
- 22022728
- Subject category
- S63: RADIATION, THERMAL, AND OTHER ENVIRONMENTAL POLLUTANT EFFECTS ON LIVING ORGANISMS AND BIOLOGICAL MATERIALS; S61: RADIATION PROTECTION AND DOSIMETRY;
- Descriptors DEI
- BEAM PROFILES; DOSE RATES; KERMA; MONTE CARLO METHOD; PHANTOMS; PHOTON BEAMS; RADIATION DOSES; SCATTERING; X-RAY DOSIMETRY
- Descriptors DEC
- BEAMS; DOSIMETRY; MOCKUP; STRUCTURAL MODELS