Published April 1, 1987 | Version v1
Journal article

A new beam position monitor for X-ray synchrotron radiation facilities

  • 1. Brookhaven National Lab., Upton, NY (USA). National Synchrotron Light Source

Description

We present here a simple device which can detect one-dimensional beam motion over a large range, with a spatial resolution of a fraction of a percent of its range, a time resolution of a few seconds, and good long-term stability. Its resolution is statistically limited, and could be improved by careful detector design. (orig.)

Additional details

Publishing Information

Journal Title
Nucl. Instrum. Methods Phys. Res., Sect. A
Journal Volume
255
Journal Issue
3
Series
Nucl. Instrum. Methods Phys. Res., Sect. A.
Journal Page Range
598-602
ISSN
0168-9002
CODEN
NIMAE

Optional Information

Contract/Grant/Project number
Contract DE-AC02-76CH00016