Published June 2016 | Version v1
Journal article

SU-F-T-121: Abdominal Compression Effectively Reduces the Interplay Effect and Enables Pencil Beam Scanning Proton Therapy of Liver Tumors

  • 1. University of Pennsylvania, Philadelphia, PA (United States)
  • 2. Universite catholique de Louvain, Brussels (Belgium)
  • 3. IBA, Louvain-la-neuve (Belgium)
  • 4. KU Leuven, Leuven (Belgium)

Description

Purpose: To study if abdominal compression can reduce breathing motion and mitigate interplay effect in pencil beam scanning proton therapy (PBSPT) treatment of liver tumors in order to better spare healthy liver volumes compared with photon therapy. Methods: Ten patients, six having large tumors initially treated with IMRT and four having small tumors treated with SBRT, were replanned for PBSPT. ITV and beam-specific PTVs based on 4D-CT were used to ensure target coverage in PBSPT. The use of an abdominal compression belt and volumetric repainting was investigated to mitigate the interplay effect between breathing motion and PBSPT dynamic delivery. An in-house Matlab script has been developed to simulate this interplay effect. The dose is computed on each phase individually by sorting all spots according to their simulated delivery timing. The final dose distribution is then obtained by accumulating all dose maps to a reference phase. Results: For equivalent target coverage PBSPT reduced average healthy liver dose by 9.5% of the prescription dose compared with IMRT/SBRT. Abdominal compression of 113.2±42.2 mmHg was effective for all 10 patients and reduced average motion by 2.25 mm. As a result, the average ITV volume decreased from 128.2% to 123.1% of CTV volume. Similarly, the average beam-specific PTV volume decreased from 193.2% to 183.3%. For 8 of the 10 patients, the average motion was reduced below 5 mm, and up to 3 repainting were sufficient to mitigate interplay. For the other two patients with larger residual motion, 4–5 repainting were needed. Conclusion: We recommend evaluation of the 4DCT motion histogram following simulation and the interplay effect following treatment planning in order to personalize the use of compression and volumetric repainting for each patient. Abdominal compression enables safe and more effective PBS treatment of liver tumors by reduction of motion and interplay effect. Kevin Souris is supported by IBA and Televie Grant from F.R.S.-FNRS. Liyong Lin is partially supported by Varian.

Additional details

Identifiers

Publishing Information

Journal Title
Medical Physics
Journal Volume
43
Journal Issue
6
Journal Page Range
vp.
ISSN
0094-2405
CODEN
MPHYA6

INIS

Country of Publication
United States
Country of Input or Organization
International Atomic Energy Agency (IAEA)
INIS RN
48090957
Subject category
S60: APPLIED LIFE SCIENCES; S61: RADIATION PROTECTION AND DOSIMETRY;
Descriptors DEI
LIVER; NEOPLASMS; PATIENTS; PROTON BEAMS; RADIATION DOSE DISTRIBUTIONS; RADIOTHERAPY; SIMULATION
Descriptors DEC
BEAMS; BODY; DIGESTIVE SYSTEM; DISEASES; GLANDS; MEDICINE; NUCLEAR MEDICINE; NUCLEON BEAMS; ORGANS; PARTICLE BEAMS; RADIOLOGY; THERAPY

Optional Information

Notes