Determination of both mechanical and electronic shifts in cone beam SPECT

  • 1. Duke Univ., Durham, NC (United States). Dept. of Radiology

Description

The difference between the displacement of the centre of rotation (mechanical shift, MS) and the electronic centring misalignment (electronic shift, ES) in cone beam SPECT is evaluated. A method is proposed to determine both MS and ES using the centroid of a projected point source sampled over 360oC and the Marquardt non-linear fitting algorithm. Both shifts are characterized by two orthogonal components. This method is verified using Monte Carlo simulated point source data with different combinations of mechanical and electronic shifts. Both shifts can be determined correctly. The proposed method was also applied to the authors' cone beam SPECT system to determine both shifts as well as the focal length. The determined ES parameters are then used to correct the projections and the MS parameters are incorporated into a reconstruction algorithm. The point source image are reconstructed and the image resolutions with and without the shift corrections are measured. (Author)

Additional details

Publishing Information

Journal Title
Physics in Medicine and Biology
Journal Volume
38
Journal Issue
6
Journal Page Range
p. 743-754.
ISSN
0031-9155
CODEN
PHMBA7

INIS

Country of Publication
United Kingdom
Country of Input or Organization
United Kingdom
INIS RN
25000689
Subject category
S62: RADIOLOGY AND NUCLEAR MEDICINE;
Descriptors DEI
ALGORITHMS; COMPARATIVE EVALUATIONS; CORRECTIONS; IMAGE PROCESSING; SINGLE PHOTON ECT; SPATIAL RESOLUTION
Descriptors DEC
COMPUTERIZED TOMOGRAPHY; EMISSION COMPUTED TOMOGRAPHY; EVALUATION; RESOLUTION; TOMOGRAPHY