Published April 1989 | Version v1
Journal article

Interactive reconstruction in single-photon tomography

  • 1. Washington Univ., St. Louis, MO (USA). Edward Mallinckrodt Inst. of Radiology

Description

A new method is described to allow interactive selection of the reconstruction filter at the time of interpretation of images from single-photon tomography. In the filtered back projection algorithm, the only part of the reconstruction process requiring user interaction is the selection of the window function. Since the ramp and window filters have different purposes, they can be separated, placing the window at the end of the reconstruction process as a three-dimensional filter. All stages of reconstruction except the window filtering are performed before the physician begins to interpret the study. The three-dimensional filtering is performed very rapidly with use of the Chebyshev convolution algorithm. A 64 x 64 x 64 pixel cube of data is filtered in 13-33 s using filters of 3-11 lengths. Smaller volumes of image data can be filtered in less than 1 s; thus, the user can interactively choose any desired filter for a given tomographic study at the time of interpretation of the images. (orig.)

Additional details

Publishing Information

Journal Title
European Journal of Nuclear Medicine
Journal Volume
15
Journal Issue
4
Series
Eur. J. Nucl. Med.
Journal Page Range
189-193
ISSN
0340-6997
CODEN
EJNMD

INIS

Country of Publication
Germany
Country of Input or Organization
Germany
INIS RN
20037335
Subject category
S62: RADIOLOGY AND NUCLEAR MEDICINE;
Descriptors DEI
FILTERS; IMAGE PROCESSING; IMAGES; SINGLE PHOTON ECT; THREE-DIMENSIONAL CALCULATIONS
Descriptors DEC
COMPUTERIZED TOMOGRAPHY; EMISSION COMPUTED TOMOGRAPHY; TOMOGRAPHY