Monte Carlo studies of x-ray scattering in transmission diagnostic radiology
Creators
- 1. Center for Radiation Research, National Bureau of Standards, Gaithersburg, Maryland 20899
Description
Monte Carlo methods have been used to simulate the scattering of x rays in polystyrene and water phantoms. In particular, the ratio of the scattered to total x-ray fluence (scatter fraction) has been calculated for monoenergetic x-ray beams in the energy region relevant to diagnostic radiology and nuclear medicine (30--660 keV). Simulations have been made for representative values of the pertinent geometrical factors; phantom thickness from 5 to 21 cm, x-ray beam diameters of 10 and 25 cm, and scatterer-to-image-plane separations from 0 to 20 cm. As a function of x-ray energy, the scatter fraction was found to vary slowly between 30 and 100 keV, and to decrease between 100 and 660 keV. The present results were generated with a special transport code which included the effects of special geometries and the response of the x-ray detector. With the inclusion of these effects, the results resolved inconsistencies and showed good agreement with previous measured and calculated data
Additional details
Publishing Information
- Journal Title
- Med. Phys.
- Journal Volume
- 13
- Journal Issue
- 4
- Series
- Med. Phys.
- Journal Page Range
- 490-495
- ISSN
- 0094-2405
- CODEN
- MPHYA
INIS
- Country of Publication
- United States
- Country of Input or Organization
- United States
- INIS RN
- 17068522
- Subject category
- S62: RADIOLOGY AND NUCLEAR MEDICINE;
- Descriptors DEI
- BIOMEDICAL RADIOGRAPHY; DIAGNOSTIC TECHNIQUES; IMAGE PROCESSING; MONTE CARLO METHOD; PHANTOMS; POLYSTYRENE; SCATTERING; WATER; X RADIATION; X-RAY RADIOGRAPHY
- Descriptors DEC
- ELECTROMAGNETIC RADIATION; HYDROGEN COMPOUNDS; IONIZING RADIATIONS; MEDICINE; MOCKUP; ORGANIC COMPOUNDS; ORGANIC POLYMERS; OXYGEN COMPOUNDS; POLAR SOLVENTS; POLYMERS; POLYOLEFINS; POLYVINYLS; RADIATIONS; SOLVENTS; STRUCTURAL MODELS