Conducting tubes threading the pick-up coils in SQUID magnetometers may generate magnetic flux noise larger than the intrinsic SQUID noise. A method for calculating the level of this noise is presented which can aid in the design of low noise systems. Application to a system in this lab predicts a noise level of 0.001 fluxoids rms/(Hz)1/sup //2, in good agreement with the observed level of 0.001 fluxoids rms in a 1-Hz bandwidth