A simple measurement method for verifying the accuracy of the displayed dose on an over-couch-type X-ray fluoroscopic system
Creators
- 1. Kanazawa University Hospital, Department of Radiology, Kanazawa, Ishikawa (Japan)
- 2. Kanazawa University, Institute of Medical, Pharmaceutical and Health Sciences, Faculty of Health Sciences, Department of Quantum Medical Technology, Kanazawa, Ishikawa (Japan)
Description
The purpose of this study was to establish a simple measurement method to verify the accuracy of incident air kerma (Ka, r) and air kerma area product (PKA) displayed on an over-couch-type X-ray fluoroscopy system. A dosimeter was located at the patient entrance reference point, and the irradiation field size was set to 10×10 cm. A lead plate was placed on the couchtop to protect the image receptor, and the duration of fluoroscopy was set to 1 min. The Ka, r was measured with the proposed method and the Japanese Industrial Standards (JIS) method on three X-ray fluoroscopy units of different manufactures. The effect of backscattered X-rays from the lead plate was calculated using Monte Carlo methods. The errors of the displayed Ka, r and PKA to the measured Ka, r and PKA with our proposed method were calculated. There was no significant difference in the measured Ka, r between the proposed method and the JIS method in all units. The effect of backscattered X-ray was ≤0.5%. The errors of displayed Ka, r and PKA to those measured were in the range of 3.4 to 15.7% and −4.1 to 20.3%, respectively, which met the tolerance for accuracy of ±35% in accordance with the JIS method. We found that our proposed method was simple and that the accuracy of measured values was comparable to that of the JIS method. (author)
Availability note (English)
Available from DOI: https://doi.org/10.6009/jjrt.2021_JSRT_77.10.1180Additional details
Additional titles
- Original title (Japanese)
- オーバーテーブル式 X 線透視装置の表示線量の簡易精度検証方法
Identifiers
Publishing Information
- Journal Title
- Nippon Hoshasen Gijutsu Gakkai Zasshi (Online)
- Journal Volume
- 77
- Journal Issue
- 10
- Series
- 雑誌名:日本放射線技術学会雑誌
- Journal Page Range
- p. 1180-1185
- ISSN
- 1881-4883
INIS
- Country of Publication
- Japan
- Country of Input or Organization
- Japan
- INIS RN
- 53096681
- Subject category
- S61: RADIATION PROTECTION AND DOSIMETRY;
- Descriptors DEI
- ACCURACY; BACKSCATTERING; DISPLAY DEVICES; DOSEMETERS; FLUOROSCOPY; KERMA; MILLI GY RANGE 01-10; P CODES; PHANTOMS; RADIATION DOSES; VERIFICATION; X RADIATION
- Descriptors DEC
- ABSORBED DOSE RANGE; BIOMEDICAL RADIOGRAPHY; COMPUTER CODES; COMPUTER OUTPUT DEVICES; COMPUTER-GRAPHICS DEVICES; DIAGNOSTIC TECHNIQUES; DOSES; ELECTROMAGNETIC RADIATION; IONIZING RADIATIONS; MEASURING INSTRUMENTS; MEDICINE; MILLI GY RANGE; MOCKUP; NUCLEAR MEDICINE; RADIATION DOSE RANGES; RADIATIONS; RADIOLOGY; SCATTERING; STRUCTURAL MODELS
Optional Information
- Notes
- 14 refs., 2 figs., 4 tabs.