Manufacture of 60Co source irradiation facility and measurement of radiation filed

  • 1. China Institute of Atomic Energy, Beijing (China)

Description

60Co source radiation facility is an important equipment to produce reference radiation filed, according to national standard, the scattered dose rate must be less than 5% of the total dose rate in the reference radiation filed. Scattered radiation contribution and uniformity of reference radiation field are important parameters to describe the radiation performance of 60CO source irradiation device, Monte Carlo method was used to study the scattered radiation and uniformity of the reference radiation filed to provide theoretical basis for the design of 60CO source irradiating device in order to avoid economic losses caused by design mistakes. When 60CO source irradiating device was manufactured, PTW ionization chamber was used to measure the irradiation field. The results showed that the scattered radiation and uniformity of the radiation filed were in good agreement with the simulation results. The radiation performance met the design requirements. (authors)

Additional details

Publishing Information

Journal Title
Journal of Isotopes
Journal Volume
26
Journal Issue
4
Journal Page Range
p. 239-243
ISSN
1000-7512

Optional Information

Notes
5 figs., 2 tabs., 20 refs.