Published 1997 | Version v1
Journal article

Application of a low-level anti-coincidence gamma-spectrometer for environmental radioactivity measurement

  • 1. Yogyakarta Nuclear Research Center (Indonesia)
  • 2. Central Dept. of Analytical Chemistry, Research Center Juelich (Germany)
  • 3. Inst. of Applied Physical Chemistry, Research Center Juelich (Germany)

Description

Sensitive environmental radioactivity measurement using direct γ-spectroscopy needs to be carried out under particular low-level conditions as the quantity to be determined generally is very close to the natural background. Passive shielding of the detectors in addition to electronic suppression of Compton background resulting from scattering of high energy γ-rays in the germanium detector seems to be the most promising approach to obtain statistically well defined peaks within reasonable measurement time. Results obtained with a heavily shielded anti-coincidence spectrometer installed at the Research Center Juelich, ideally suited for the measurement of low-level environmental radioactivity in various materials are described. (author)

Additional details

Publishing Information

Journal Title
Journal of Radioanalytical and Nuclear Chemistry
Journal Volume
223
Journal Issue
1-2
Journal Page Range
p. 177-185.
ISSN
0236-5731
CODEN
JRNCDM

Optional Information

Notes
17 refs.