Photodensitometer measurements of images on radiographs
Description
An x-ray photodensitometer provided accurate and precise measurements of linear dimensions on radiograph images. These measurements helped to infer the hidden dimensions of objects radiographed. A goal was to achieve an accuracy of +-25 μm. The recorded accuracy did not measure up to the precision, which, under favorable conditions, was +-25 μm. Systematic errors caused the errors in absolute value to be considerably higher in some cases. To satisfy this, auxiliary calibrations were made. Many deterrents to successful measurement arose. These included film grain noise, excessive density range and edge penetration, blurring, interfering structures, and scattered radiation. Film grain noise, in particular, limited the resolution of small details (fine structure) in a radiograph. Because of so many deterrents, the applicability of the measurement method used appears quite limited
Availability note (English)
MF available from INIS under the Report Number; Available from NTIS.Files
7268989.pdf
Files
(436.5 kB)
| Name | Size | Download all |
|---|---|---|
|
md5:6cfd546dd0a907c7bf213233ebed927d
|
436.5 kB | Preview Download |
Additional details
Publishing Information
- Imprint Pagination
- 25 p.
- Report number
- UCRL--52037
INIS
- Country of Publication
- United States
- Country of Input or Organization
- United States
- INIS RN
- 7268989
- Subject category
- S42: ENGINEERING; S46: INSTRUMENTATION RELATED TO NUCLEAR SCIENCE AND TECHNOLOGY;
- Descriptors DEI
- AUTOMATION; DATA PROCESSING; DENSITOMETERS; IMAGES; PHOTOGRAPHIC FILMS; X-RAY RADIOGRAPHY
- Descriptors DEC
- MEASURING INSTRUMENTS; PHOTOMETERS
Optional Information
- Notes
- Available from NTIS. $4.00.