Size-specific dose estimation in the Korean lung cancer screening project: Dose a 32 cm diameter phantom represent a standard-sized patient in Korean population?
Creators
- 1. Gachon University Gil Medical Center, Incheon (Korea, Republic of)
- 2. Samsung Medical Center, Sungkyunkwan University School of Medicine, Seoul (Korea, Republic of)
- 3. Seoul National University College of Medicine, Seoul (Korea, Republic of)
- 4. National Cancer Center, Goyang (Korea, Republic of)
- 5. Pusan National University Hospital, Busan (Korea, Republic of)
- 6. Chungbuk National University Hospital, Cheongju (Korea, Republic of)
Description
The purposes of this study were to evaluate size-specific dose estimate (SSDE) of low-dose CT (LDCT) in the Korean Lung Cancer Screening (K-LUCAS) project and to determine whether CT protocols from Western countries are appropriate for lung cancer screening in Korea. For participants (n = 256, four institutions) of K-LUCAS pilot study, volume CT dose index (CTDIvol) using a 32-cm diameter reference phantom was compared with SSDE, which was recalculated from CTDIvol using size-dependent conversion factor (f-size) based on the body size, as described in the American Association of Physicists in Medicine Report 204. This comparison was subsequently assessed by body mass index (BMI) levels (underweight/normal vs. overweight/obese), and automatic exposure control (AEC) adaptation (yes/no). Size-specific dose estimate was higher than CTDIvol (2.22 ± 0.75 mGy vs. 1.67 ± 0.60 mGy, p < 0.001), since the f-size was larger than 1.0 for all participants. The ratio of SSDE to CTDIvol was higher in lower BMI groups; 1.26, 1.37, 1.43, and 1.53 in the obese (n = 103), overweight (n = 70), normal (n = 75), and underweight (n = 4), respectively. The ratio of SSDE to CTDIvol was greater in standard-sized participants than in large-sized participants independent of AEC adaptation; with AEC, SSDE/CTDIvol in large- vs. standard-sized participants: 1.30 ± 0.08 vs. 1.44 ± 0.08 (p < 0.001) and without AEC, 1.32 ± 0.08 vs. 1.42 ± 0.06 (p < 0.001). Volume CT dose index based on a reference phantom underestimates radiation exposure of LDCT in standard-sized Korean participants. The optimal radiation dose limit needs to be verified for standard-sized Korean participants
Additional details
Publishing Information
- Journal Title
- Korean Journal of Radiology
- Journal Volume
- 19
- Journal Issue
- 6
- Series
- 12 refs, 3 figs, 3 tabs
- Journal Page Range
- p. 1179-1186
- ISSN
- 1229-6929
INIS
- Country of Publication
- Korea, Republic of
- Country of Input or Organization
- Korea, Republic of
- INIS RN
- 50047773
- Subject category
- S62: RADIOLOGY AND NUCLEAR MEDICINE;
- Descriptors DEI
- LUNGS; NEOPLASMS; PATIENTS; POPULATIONS; RADIATION DOSES; REPUBLIC OF KOREA; VERIFICATION
- Descriptors DEC
- ASIA; BODY; DEVELOPING COUNTRIES; DISEASES; DOSES; ORGANS; RESPIRATORY SYSTEM