Published August 2015 | Version v1
Miscellaneous

Precision tests with Leica lasertracker AT402 in short-range measurement

  • 1. SPring-8 Service Co., Ltd., Tatsuno, Hyogo (Japan)
  • 2. RIKEN, Sayo, Hyogo (Japan)

Description

We have carried out precision tests with Leica lasertracker AT402 in short-range (4-6m) measurement. Its typical precision value is 7.5μm+3μm/m. We set 6 measured points on stone precision tables, and obtained 24 data sets with different tracker positions. From the measurement, deviations (1σ) of the three-dimensional precision were 9μm for 4m, and 13μm for 6m, and the total precision was estimated to be 1.3μm+2μm/m, Using this value, a calculated error (1σ) of network analysis was estimated to be 8μm and it was good agreement with obtained deviation of the most probable values by network analysis using 8 data sets with 3 tracker positions. (author)

Part of:
Proceedings of the 12th annual meeting of Particle Accelerator Society of Japan

Additional details

Publishing Information

Imprint Title
Proceedings of the 12th annual meeting of Particle Accelerator Society of Japan
Imprint Pagination
1419 p.
Journal Page Range
p. 1384-1388

Conference

Title
12. annual meeting of Particle Accelerator Society of Japan
Dates
5-7 Aug 2015
Place
Tsuruga, Fukui (Japan)

INIS

Country of Publication
Japan
Country of Input or Organization
Japan
INIS RN
48024980
Subject category
S43: PARTICLE ACCELERATORS;
Resource subtype / Literary indicator
Conference, Non-conventional Literature
Descriptors DEI
ACCURACY; ALIGNMENT; COLLIMATORS; COMPUTER CODES; DATA COVARIANCES; ERRORS; FREE ELECTRON LASERS; LASER TARGETS; NETWORK ANALYSIS; WIGGLER MAGNETS; X-RAY LASERS
Descriptors DEC
EQUIPMENT; LASERS; MAGNETS; TARGETS

Optional Information

Notes
2 refs., 9 figs., 4 tabs.