Published December 1, 2015 | Version v1
Journal article

Solid scintillator based neutron fluctuation measurement on EAST tokamak

  • 1. Institute of Plasma Physics, Chinese Academy of Sciences, Hefei 230031 (China)
  • 2. Department of Physics and Astronomy, University of California, Irvine, California 92697-4575 (United States)

Description

Microsecond level fast temporal resolved neutron flux and its fluctuation measurement system based on three types of solid scintillator detectors has been successfully established on the Experimental Advanced Superconducting Tokamak (EAST) for energetic particle (EP) and magnetohydrodynamics (MHD) instabilities relevant studies. The detector #1, where 50mm thick polyethylene is used for neutron thermalization, is mostly sensitive to thermal neutron. The detector #2 and #3 measure fast D-D neutrons directly with different gamma immunity. Design details together with detector test results with three types of radioisotope sources are presented. The system has been successfully implemented in EAST experiments for neutron and gamma identification. Typical fast MHD fluctuation related EAST experimental results from this system is also presented

Availability note (English)

Available from http://dx.doi.org/10.1088/1748-0221/10/12/P12013

Additional details

Publishing Information

Journal Title
Journal of Instrumentation
Journal Volume
10
Journal Issue
12
Journal Page Range
p. P12013
ISSN
1748-0221