Published 2006 | Version v1
Journal article

Glow peak temperature, supra-linearity and LET dependence of TLDS - Correlation studies

  • 1. Health, Safety and Environment Group, Bhabha Atomic Research Centre, Mumbai 400085 (India)

Description

The study of the well separated low and high temperature glow peaks in CaSO4:Dy and CaF2:Tm was undertaken to develop comparative data on supra-linearity of low and high temperature glow peaks for their use in the estimation of elapsed time of exposure and to analyse their responses to low and high-linear energy transfer (LET) radiation. In CaSO4:Dy, unlike its dosimetric peak, the structure of glow peaks at ∼140 and 400 deg. C remains unchanged (peak position changes within 4 deg. C) with 60Co gamma-ray exposure up to 1 kGy air kerma. The glow peaks at 140 deg. C exhibited higher supra-linearity than that of the peaks at 240 and 400 deg. C. In CaF2:Tm, 110 deg. C glow peak exhibited higher supra-linearity and higher response to high-LET radiation as compared with 150 deg. C glow peak. No correlation between glow peak temperature and supra-linearity or the LET response was observed. (authors)

Availability note (English)

Available from doi: http://dx.doi.org/10.1093/rpd/nci559

Additional details

Identifiers

Publishing Information

Journal Title
Radiation Protection Dosimetry
Journal Volume
119
Journal Issue
1-4
Journal Page Range
p. 276-279
ISSN
0144-8420

Conference

Title
14. International Conference on Solid State Dosimetry
Acronym
SSD14
Dates
27 Jun - 2 Jul 2004
Place
New Haven (United States)

Optional Information

Notes
26 refs