Published 2006 | Version v1
Journal article

The implementation in routine of the ENEA new personal photon dosemeter

  • 1. ENEA, Radiation Protection Inst., via dei Colli, 16, 40136 Bologna (Italy)
  • 2. DIENCA, Dept. of Energy, Nuclear and Environmental Engineering, Univ. of Bologna, via dei Colli, 16, 40136 Bologna (Italy)

Description

The ENEA photon dosemeter, introduced in 1995, consisting of two differently filtrated LiF(Mg,Cu,P) detectors, has been modified recently. The ABS (acrylonitrile butadiene styrene) plastic support has been replaced by a new aluminium card supporting the same two detectors (LiF(Mg,Cu,P) GR200). The new card, fully developed at the ENEA-Radiation Protection Inst. (which is going to be patented), can now be processed through a Harshaw Model 6600 Automated TLD Reader, a hot gas reader. This paper reports the results of the individual calibration of ∼60,000 LiF(Mg,Cu,P) GR200 detectors inserted on the new aluminium cards. Before the implementation in routine of the new cards, the reader has been characterised. Steps and tests to be made to use the card in routine (i.e. reader stability, linearity, reproducibility, etc.) are reported. The whole dosimetric system now combines the very good performances of the Harshaw Model 6600 reader and that of LiF(Mg,Cu,P) thermoluminescent material. (authors)

Availability note (English)

Available from doi: http://dx.doi.org/10.1093/rpd/nci565

Additional details

Identifiers

Publishing Information

Journal Title
Radiation Protection Dosimetry
Journal Volume
120
Journal Issue
1-4
Journal Page Range
p. 278-282
ISSN
0144-8420

Conference

Title
14. International Conference on Solid State Dosimetry
Acronym
SSD14
Dates
27 Jun - 2 Jul 2004
Place
New Haven (United States)

Optional Information

Notes
5 refs