Published March 1992 | Version v1
Journal article

Moisture monitors for the tritium emissions reduction facility

  • 1. EG and G Mound Applied Technologies, Miamisburg, OH (United States)

Description

This paper reports on three types of moisture monitors: MCM Model Dewluxe-20, Panametrics Model System One, and Shaw Model SHA-TRSD, tested for accuracy and speed of response over low (10-50 ppm H2O), medium (100-500 ppm H2O), and high (500-4,000 ppm H2O) concentration ranges. The results for the three instruments tested (one of each kind) showed that the MCM instrument was generally more accurate and responded more quickly than the other two instruments, with the Panametrics instrument being less accurate (except at low concentration) and slower to respond, and the Shaw instrument was the least accurate and least responsive of the three instruments during the tests made

Additional details

Publishing Information

Journal Title
Fusion Technology
Journal Volume
21
Journal Issue
2
Journal Page Range
p. 452-456.
ISSN
0748-1896
CODEN
FUSTE8

Conference

Title
4. topical meeting on tritium technology in fission, fusion, and isotopic applications.
Dates
29 Sep - 4 Oct 1991.
Place
Albuquerque, NM (United States).

Optional Information

Secondary number(s)
CONF-910920--.