Published February 21, 2007 | Version v1
Journal article

Computer-aided detection system for clustered microcalcifications: comparison of performance on full-field digital mammograms and digitized screen-film mammograms

  • 1. Department of Radiology, University of Michigan, CGC B2103, 1500 E Medical Center Drive, Ann Arbor, MI 48109 (United States)

Description

We have developed a computer-aided detection (CAD) system to detect clustered microcalcifications automatically on full-field digital mammograms (FFDMs) and a CAD system for screen-film mammograms (SFMs). The two systems used the same computer vision algorithms but their false positive (FP) classifiers were trained separately with sample images of each modality. In this study, we compared the performance of the CAD systems for detection of clustered microcalcifications on pairs of FFDM and SFM obtained from the same patient. For case-based performance evaluation, the FFDM CAD system achieved detection sensitivities of 70%, 80% and 90% at an average FP cluster rate of 0.07, 0.16 and 0.63 per image, compared with an average FP cluster rate of 0.15, 0.38 and 2.02 per image for the SFM CAD system. The difference was statistically significant with the alternative free-response receiver operating characteristic (AFROC) analysis. When evaluated on data sets negative for microcalcification clusters, the average FP cluster rates of the FFDM CAD system were 0.04, 0.11 and 0.33 per image at detection sensitivity level of 70%, 80% and 90% compared with an average FP cluster rate of 0.08, 0.14 and 0.50 per image for the SFM CAD system. When evaluated for malignant cases only, the difference of the performance of the two CAD systems was not statistically significant with AFROC analysis

Additional details

Identifiers

DOI
10.1088/0031-9155/52/4/008;
PII
S0031-9155(07)28596-3;

Publishing Information

Journal Title
Physics in Medicine and Biology
Journal Volume
52
Journal Issue
4
Journal Page Range
p. 981-1000
ISSN
0031-9155
CODEN
PHMBA7

INIS

Country of Publication
United Kingdom
Country of Input or Organization
International Atomic Energy Agency (IAEA)
INIS RN
38072707
Subject category
S62: RADIOLOGY AND NUCLEAR MEDICINE;
Descriptors DEI
ALGORITHMS; COMPARATIVE EVALUATIONS; DETECTION; DIAGNOSTIC TECHNIQUES; DIAGNOSTIC USES; IMAGES; PATIENTS; PERFORMANCE; SENSITIVITY
Descriptors DEC
EVALUATION; MATHEMATICAL LOGIC; USES