Published October 2011 | Version v1
Journal article

Calibration of activity readings from radiopharmaceuticals with impurities

  • 1. Institute of Radiological Science, Tzi-Chi College of Technology, Hualien, Taiwan (China). Department of Radiological Technology
  • 2. Institute of Nuclear Energy Research, Lung-Tan, Taiwan (China). Division of Isotope

Description

Accurately determining the amount of a radionuclide actually administered to a patient is an important aspect of quality control in nuclear medicine. In this study, in order to accurately assess the 'apparent' radioactivities of radionuclides with impurities by a dose calibrator, two methods, (1) curve fitting, and (2) relative sensitivities calculation, have been employed to determine the parameter, M (relative response). From the comparisons between the measurement and re-calculated radioactivity using the M values, each method was found to have its own advantage and disadvantage, depending upon the fraction ratio of the impurity as well as the energy level of the impurity's photons. This study might be helpful for improving the assessment of radiopharmaceuticals' 'true' activity and its 'apparent' activity measured by a dose calibrator. (author)

Additional details

Publishing Information

Journal Title
Journal of Radioanalytical and Nuclear Chemistry
Journal Volume
290
Journal Issue
1
Journal Page Range
p. 153-157
ISSN
0236-5731
CODEN
JRNCDM

Optional Information

Notes
9 refs.