Neutron spin echo device to improve the energy resolution of triple axis spectrometers
Description
The design of a spin-echo option to the D10 thermal neutron Triple axis spectrometer at I.L.L. is described. The option consists of vertically focusing Heusler crystal polarizer/monochromator and analyzer systems, transverse field electromagnets for the Larmor processions and the necessary spin turn devices. The polarized neutron intensity at the sample position is reduced by a factor of 5 with respect to the standard using P.G. crystals. The resolution obtained is 1 μeV at 14 meV incident energy
Additional details
Publishing Information
- Journal Title
- AIP Conf. Proc.
- Journal Issue
- no.89
- Series
- AIP Conf. Proc.
- ISSN
- 0094-243X
- CODEN
- APCPC
INIS
- Country of Publication
- United States
- Country of Input or Organization
- United States
- INIS RN
- 17028608
- Subject category
- S46: INSTRUMENTATION RELATED TO NUCLEAR SCIENCE AND TECHNOLOGY;
- Resource subtype / Literary indicator
- Numerical Data
- Descriptors DEI
- ENERGY RESOLUTION; EXPERIMENTAL DATA; MONOCHROMATORS; NEUTRON SPECTROMETERS; POLARIZATION; SPECIFICATIONS; SPIN ECHO; SUPERCONDUCTING MAGNETS
- Descriptors DEC
- DATA; ELECTRICAL EQUIPMENT; ELECTROMAGNETS; EQUIPMENT; INFORMATION; MAGNETS; MEASURING INSTRUMENTS; NUMERICAL DATA; RESOLUTION; SPECTROMETERS; SUPERCONDUCTING DEVICES