LC circuit as a pulse-height analyser for NaI gamma spectrometry

  • 1. Department of Nuclear Engineering, University of Sofia, 5, J. Bourchier Blvd., BG-1164 Sofia (Bulgaria)
  • 2. Department of Electronic Technique, Technical University of Sofia, 8, Kl.Ohridski Blvd., BG-1000 Sofia (Bulgaria)
  • 3. Institute for Nuclear Research and Nuclear Energy, Bulgarian Academy of Science, 72, Tzarigradsko chaussee Blvd., BG-1784 Sofia (Bulgaria)

Description

An LC circuit used as anode load of a photomultiplier tube converts its output current pulse into a damped harmonically oscillating voltage. Its amplitude is not only proportional to the transferred charge but it can be also readily encoded by the count of the successive maxima it passes until it falls below a predefined voltage level. Thus, a scintillation spectrometer may be realised in a very simple way, bypassing the classical (but rather sophisticated) solutions based on pulse-shaping amplifiers and amplitude-to-digital converters. The structure, spectral response and some experiments with a 256-channel 'LC spectrometer' are discussed. A 7% energy resolution has been achieved at 662 keV (with a diameter 45x40 mm NaI(Tl) crystal). Despite some uncommon features (arising from the logarithmic channel-energy dependence), this spectrometer has useful potential in at least two fields of scintillation gamma-spectroscopy - as a portable dose-rate meter or in the study of the soft component of the cosmic rays

Additional details

Identifiers

DOI
10.1016/j.nima.2004.06.110;
PII
S0168-9002(04)01358-0;

Publishing Information

Journal Title
Nuclear Instruments and Methods in Physics Research. Section A, Accelerators, Spectrometers, Detectors and Associated Equipment
Journal Volume
533
Journal Issue
3
Journal Page Range
p. 509-515
ISSN
0168-9002
CODEN
NIMAER

Optional Information

Copyright
Copyright (c) 2004 Elsevier Science B.V., Amsterdam, The Netherlands, All rights reserved.