A calculation method to determine γ-ray peak efficiency was developed, which was based on numerical integration of peak efficiency calculated by γ-ray path length through a Ge detector. The calculation needs some physical parameters of a Ge detector, such as crystal size and dead layer thickness, sample size and elemental composition of sample matrix, and one or two experimentally obtained peak efficiency for normalization. The method determines peak efficiency including self-absorption correction for a volume source consisted of any kind of matrix. (author)
Additional details
Publishing Information
- Journal Title
- Radioisotopes (Tokyo)
- Journal Volume
- 48
- Journal Issue
- 10
- Journal Page Range
- p. 617-625
- ISSN
- 0033-8303
INIS
- Country of Publication
- Japan
- Country of Input or Organization
- Japan
- INIS RN
- 31005712
- Subject category
- S46: INSTRUMENTATION RELATED TO NUCLEAR SCIENCE AND TECHNOLOGY;
- Descriptors DEI
- CALCULATION METHODS; CESIUM 137; COMPUTER CALCULATIONS; GAMMA DETECTION; GAMMA SOURCES; GE SEMICONDUCTOR DETECTORS; GEOMETRY; PEAKS; SELF-ABSORPTION
- Descriptors DEC
- ABSORPTION; BETA DECAY RADIOISOTOPES; BETA-MINUS DECAY RADIOISOTOPES; CESIUM ISOTOPES; DETECTION; INTERMEDIATE MASS NUCLEI; ISOTOPES; MATHEMATICS; MEASURING INSTRUMENTS; NUCLEI; ODD-EVEN NUCLEI; RADIATION DETECTION; RADIATION DETECTORS; RADIATION SOURCES; RADIOISOTOPES; SEMICONDUCTOR DETECTORS; SORPTION; YEARS LIVING RADIOISOTOPES