Published October 1980
| Version v1
Report
Open
Low noise PWC cathode readout system
Description
A system has been developed, primarily to detect the induced charge deposited on PWC cathodes, which is versatile, fast and has a good signal to noise ratio for signals of greater than or equal to 10-14 Coulomb input. The amplifier system, which is completely separated from the detector by 95 Ω coaxial cables, is followed by a new charge integrating, version of the SHAM/BADC system developed at SLAC. This SHAM IV system is CAMAC based, allowing for computer calibration of the entire system from amplifier through ADC
Availability note (English)
MF available from INIS under the Report Number; Available from NTIS., PC A02/MF A01.Files
12592763.pdf
Files
(191.1 kB)
| Name | Size | Download all |
|---|---|---|
|
md5:1dd0621e9bd6d42d3d5dc125d932fe7d
|
191.1 kB | Preview Download |
Additional details
Publishing Information
- Imprint Pagination
- 6 p.
- Report number
- SLAC-PUB--2641
Conference
- Title
- IEEE nuclear science symposium.
- Dates
- 5 - 7 Nov 1980.
- Place
- Orlando, FL, USA.
INIS
- Country of Publication
- United States
- Country of Input or Organization
- United States
- INIS RN
- 12592763
- Subject category
- S46: INSTRUMENTATION RELATED TO NUCLEAR SCIENCE AND TECHNOLOGY;
- Resource subtype / Literary indicator
- Conference
- Descriptors DEI
- AMPLIFIERS; CALIBRATION; CAMAC SYSTEM; CATHODES; PROPORTIONAL COUNTERS; READOUT SYSTEMS
- Descriptors DEC
- ELECTRODES; MEASURING INSTRUMENTS; RADIATION DETECTORS
Optional Information
- Secondary number(s)
- CONF-801103--30.