Published June 2002 | Version v1
Journal article

Calibration of 1220 Quantulus liquid scintillation spectrometer for tritium counting efficiency

Creators

  • 1. Zhejiang Radiation Environmental Monitoring Center, Hangzhou (China)

Description

This paper described the method of calibrating 1220 Quantulus Liquid Scintillation spectrometer for tritium counting efficiency. The results show that the commercially available quenching correction standard source is used for calibration purpose, its adaptability to the samples to be measured should be checked at the first. Otherwise, there will be larger error in the measuring results of samples. For the routine measurement of tritium, the same preparation procedures can be used, with quenching efficient SQP (E) within the range of 711 to 732. Thus, quenching correction source series can be prepared with tritium-free water as quenching materials. On this basis, the quenching correction curve is obtained to calibrate the instrument's counting efficiency. (author)

Additional details

Publishing Information

Journal Title
Radiation Protection Bulletin
Journal Volume
22
Journal Issue
3
Journal Page Range
p. 15-18
ISSN
1004-6356

Optional Information

Notes
6 tabs., 2 refs.