Published 1988 | Version v1
Book

Influence of pre- and postprocessing filters on the image quality of Tc-99m, Ga-67, and TI-201 SPECT

Description

The authors evaluated preprocessing filters to determine their effect on the resolution, contrast, and lesion detectibility of single photon emission. Liver and heart phantom data were acquired with Tc-99m or T1-201. Twelve patient studies with Tc-99m, Ga-67, and T1-201 were evaluated. Data were preprocessed with a nine-point weighted, nonweighted, Butterworth, or no smoothing filter. Preprocessing reduced contrast up to 30% in cold and 10% in hot lesions. In patients, preprocessing did not affect hot lesions, but cold lesions could be obscured. Postprocessing reduced motion artifacts. In conclusion, preprocessing is unnecessary; a proper reconstruction filter is adequate for noise suppression

Additional details

Publishing Information

Publisher
Radiological Society of North America Inc.
Imprint Place
Oak Brook, IL (USA)
Imprint Title
Radiological Society of North America 74th scientific assembly and annual meeting (Abstracts)
Imprint Pagination
395 p.
Journal Page Range
p. 214.

Conference

Title
74. scientific assembly and annual meeting of the Radiological Society of North America (RSNA).
Dates
27 Nov - 2 Dec 1988.
Place
Chicago, IL (USA).

Optional Information

Secondary number(s)
CONF-8811134--.