Description
A new anode support structure for the spherical proportional counter is presented that incorporates a resistive correction electrode made of glass. This electrode improves the electric field homogeneity versus angle, while suppressing the probability and intensity of sparks compared to non-resistive alternatives. The configuration of the correction electrode was optimised with simulations. Such support structures have been constructed and measurements have demonstrated homogeneous response of the detector and operational stability. A measurement of the resistivity of the glass used is also presented.
Availability note (English)
Available from http://dx.doi.org/10.1088/1748-0221/13/11/P11006Additional details
Identifiers
Publishing Information
- Journal Title
- Journal of Instrumentation
- Journal Volume
- 13
- Journal Issue
- 11
- Journal Page Range
- p. P11006
- ISSN
- 1748-0221
INIS
- Country of Publication
- United Kingdom
- Country of Input or Organization
- International Atomic Energy Agency (IAEA)
- INIS RN
- 51051180
- Subject category
- S46: INSTRUMENTATION RELATED TO NUCLEAR SCIENCE AND TECHNOLOGY;
- Descriptors DEI
- COMPARATIVE EVALUATIONS; COMPUTERIZED SIMULATION; ELECTRIC FIELDS; GLASS; PROPORTIONAL COUNTERS; SPHERICAL CONFIGURATION; STABILITY
- Descriptors DEC
- CONFIGURATION; EVALUATION; MEASURING INSTRUMENTS; RADIATION DETECTORS; SIMULATION