CT angiography of the neck: Value of contrast medium dose reduction with low tube voltage and high tube current in a 64–detector row CT
Creators
- 1. Clinical Medical College, Yangzhou University, Radiology Department, Subei People's Hospital of Jiangsu Province, Yangzhou (China)
- 2. Radiology Department, Huashan Hospital, Fudan University, Shanghai (China)
Description
Aim: To evaluate the feasibility of a low-dose contrast medium protocol for 64-detector row computed tomography angiography (CTA) of the neck using a low-tube-voltage/high-tube-current setting. Materials and methods: A phantom study was performed using 64-detector row spiral CT at multiple tube voltage and current settings. Iodine contrast medium attenuation curves were acquired by processing and used to select the best contrast medium-to-noise ratio (CNR). A prospective clinical study was then performed on 84 patients requiring neck CTA. Patients were randomly divided into two groups of 42. Group A was examined using the conventional imaging protocol (120 kV, 400 mAs) and group B was examined at 80 kV and 600 mAs along with a 50% reduction in contrast medium dose. The CT dose index-volume (CTDIvol), background noise (BN), and CNR were measured and statistically analysed. Various image quality criteria were evaluated by two senior radiologists using a qualitative five-point scale. Results: Comparing group B with A, CTDIvol decreased by 54% (B: 27.48 mGy, A: 59.11 mGy), however, the CNR increased by 50%. The mean attenuation, which was caused by venous streak artefacts, was significantly lower in group B than A. Qualitative image analysis found that all criteria were significantly better for group B than A. Conclusion: At 64-detector row spiral CT, the low-tube-voltage/high-tube-current with low-dose contrast medium protocol was superior to the conventional protocol regarding radiation dose, venous streak artefacts, and image quality, and is feasible for CTA of the neck
Availability note (English)
Available from http://dx.doi.org/10.1016/j.crad.2013.12.001Additional details
Identifiers
- DOI
- 10.1016/j.crad.2013.12.001;
- PII
- S0009-9260(13)00570-9;
Publishing Information
- Journal Title
- Clinical Radiology
- Journal Volume
- 69
- Journal Issue
- 4
- Journal Page Range
- p. e183-e189
- ISSN
- 0009-9260
- CODEN
- CLRAAG
INIS
- Country of Publication
- United Kingdom
- Country of Input or Organization
- International Atomic Energy Agency (IAEA)
- INIS RN
- 46120105
- Subject category
- S62: RADIOLOGY AND NUCLEAR MEDICINE;
- Descriptors DEI
- BACKGROUND NOISE; BIOMEDICAL RADIOGRAPHY; BLOOD VESSELS; COMPUTERIZED TOMOGRAPHY; CONTRAST MEDIA; IMAGE PROCESSING; NECK; RADIATION DOSES
- Descriptors DEC
- BODY; CARDIOVASCULAR SYSTEM; DIAGNOSTIC TECHNIQUES; DOSES; MEDICINE; NOISE; NUCLEAR MEDICINE; ORGANS; PROCESSING; RADIOLOGY; TOMOGRAPHY
Optional Information
- Copyright
- Copyright (c) 2013 Elsevier Science B.V., Amsterdam, The Netherlands, All rights reserved.